Technical data
11.5 Cable requirements
SITRANS F M MAG 5000/6000
72 Operating Instructions, 12/2013, A5E02338368-002
11.5 Cable requirements
Coil
cable
Electrode
cable
Basic data
No. of conductors
2
3
Min. sqr. area
0.5 mm
2
0.2 mm
2
Screen
Yes
Yes
Max. capacitance
N/A
350 pF/m
Max. cable loop resistance
Media temperature:
< 100 °C
40 Ω
N/A
> 200 °C
6 Ω
N/A
Cable glands on sensor
M20x1.5 gland - Cable ø 5 ... 13 mm (0.20 ... 0.51 inches)
½ NPT gland - cable ø 5 ... 9 mm (0.20 ... 0.35 inches)
WARNING
Cable glands
For Ex zone 1 installations only certified cable glands with protection type "e" can be used
for the power supply and the coil cable. The cable glands must be approved for the actual
temperature and cable dimension.