Alarm, error, and system messages
8.2 List of error numbers
SITRANS F M MAG 5000/6000
56 Operating Instructions, 12/2013, A5E02338368-002
P42 - Current output
Check cables
Current loop is disconnected or the loop resistance
is too big
P43 - Internal error
Switch off and on
Too many errors occurred at the same time.
Some errors are not detected correctly
SENSORPROM
®
unit has been used as CT version
F60 - CAN comm. error
Transmitter/AOM
CAN bus communication error. An add-on module,
the display module or the transmitter is defective
F60 - SENSORPROM® error
Replace
It is not possible to rely on the data in
SENSORPROM
®
unit anymore
F62 - SENSORPROM® ID
Replace
The SENSORPROM
®
unit ID does not comply with
the product ID.
The SENSORPROM
®
unit is from another type of
product SITRANS F C, SITRANS F US etc.
F63 - SENSORPROM®
Replace
It is not possible to read from the SENSORPROM
®
unit anymore
F70 - Coil current
Check cables
Coil excitation has failed
F71 - Internal error
Replace transmitter
Internal convertion error in ASIC