Alarm, error, and system messages
8.2 List of error numbers
SITRANS F M MAG 5000/6000
56 Operating Instructions, 12/2013, A5E02338368-002
Error
No.
Error text
Remedy text
Comment
Output
status
Input
status
42
P42 - Current output
Check cables
Current loop is disconnected or the loop resistance
is too big
Active
Active
43
P43 - Internal error
Switch off and on
Too many errors occurred at the same time.
Some errors are not detected correctly
Active
Active
44
P44 - CT SENSORPROM®
SENSORPROM
®
unit has been used as CT version
Active
Active
60
F60 - CAN comm. error
Transmitter/AOM
CAN bus communication error. An add-on module,
the display module or the transmitter is defective
Zero
Inactiv
e
61
F60 - SENSORPROM® error
Replace
It is not possible to rely on the data in
SENSORPROM
®
unit anymore
Active
Active
62
F62 - SENSORPROM® ID
Replace
The SENSORPROM
®
unit ID does not comply with
the product ID.
The SENSORPROM
®
unit is from another type of
product SITRANS F C, SITRANS F US etc.
Zero
Inactiv
e
63
F63 - SENSORPROM®
Replace
It is not possible to read from the SENSORPROM
®
unit anymore
Active
Active
70
F70 - Coil current
Check cables
Coil excitation has failed
Active
Active
71
F71 - Internal error
Replace transmitter
Internal convertion error in ASIC
Active
Active