Technical data
11.2 Accuracy
SITRANS F M MAG 5000/6000
Operating Instructions, 12/2013, A5E02338368-002 67
Figure 11-2 MAG 6000 with MAG 1100 (not PFA), MAG 1100 F (not PFA), MAG 5100 W, MAG 3100 and MAG 3100 P
Reference conditions
(ISO 9104 and DIN/EN 29104)
A calibration certificate is shipped with every sensor and calibration data are stored in
SENSORPROM memory unit.
Medium temperature
20°C ± 5°C (68°F ± 9°C)
Ambient temperature
20°C ± 5°C (68°F ± 9°C)
Supply voltage
U
n
± 1%
Warming-up time
30 minutes
Incorporation in conductive pipe section
Inlet section
Outlet section
10 x DN (DN ≤ 1200/48")
5 x DN (DN > 1200/48")
5 x DN (DN ≤ 1200/48")
3 x DN (DN > 1200/48")
Flow conditions
Developed flow profile
Reference conditions for sensor calibration