Modbus RTU RS-485
21
SFIDK.PS.023.D2.02
6. Modbus holding registers
Modbus Modbus No. of Data Read/
register address bytes type
Description
write
4:02301 2300 2 Byte
Input Function:
0 = Off
1 = Start Batch
2 = Hold/Continue
3 = Stop Batch
4 = Zero Adjust R/W
5 = Totalizer reset
6 = Force output
7 = Freeze output
8 = Qmax 2 (night)
9 = Manual cleaning
4:02302 2301 2 Byte
Totalizer reset:
0 = Totalizer 1
R/W
1 = Totalizer 2
2 = Totalizer 1 + 2
4: 02303 2302 4 Float Force output (1/100%) R/W
MAG 6000
Modbus Modbus No. of Data Read/
register address bytes type
Description
write
4:02301 2300 2 Byte
Input selection:
0 = Off
1 = Start Batch
2 = Hold/Continue
3 = Stop Batch R/W
4 = Zero Adjust
5 = Totalizer reset
6 = Force output
7 = Freeze output
4:02302 2301 2 Byte
Totalizer reset:
0 = Totalizer 1
R/W
1 = Totalizer 2
2 = Totalizer 1 + 2
4: 02303 2302 4 Float Force output (1/100%) R/W
MASS 6000
Modbus Modbus No. of Data Read/
register address bytes type
Description
write
4:02401 2400 4 Float Sensor size (m)
4:02403 2402 4 Float Calibration factor
4:02405 2404 4 Float Correction factor
4:02421 2420 2 Byte
Excitation frequency:
0 = 3.125 Hz (50 Hz)
1 = 6.25 Hz (50 Hz)
2 = 12.5 Hz (50 Hz)
3 = 25 Hz (50 Hz)
4 = 1.5625 Hz (50 Hz)
5 = 44 Hz (only SV)
6 = 2.0833 Hz
7 = 1.875 Hz (60 Hz)
8 = 3.75 Hz (60 Hz)
9 = 7.5 Hz (60 Hz)
10 = 15 Hz (60 Hz)
11 = 30 Hz (60 Hz)
MAG 6000
Modbus Modbus No. of Data Read/
register address bytes type
Description
write
4:02401 2400 4 Float Sensor size (m) R/W
4:02403 2402 4 Float Calibration factor (us2/kg) R/W
4:02405 2404 4 Float Correction factor R/W
4:02407 2406 4 Float Sensor TC (1/100% / °C) R/W
4:02409 2408 4 Float Density parm. A (kg/m
3
) R/W
4:02411 2410 4 Float Density parm. B (*10E6) R/W
4:02413 2412 4 Float Density TC (1/100% / °C) R/W
4:02415 2414 4 Float Density offset (kg/m
3
) R/W
4:02417 2416 4 Float Density factor R/W
4:02419 2418 4 Float Table slope (kg/s) R/W
4:02422 2421 4 Float Fraction offset (kg/s) R/W
MASS 6000
6.7 External input
6.8 Sensor characteris-
tics