Index Parameter name Data type
[size]
Access Annotation
32 IM_Revision_Coun‐
ter
Unsigned16 Read only The device has parameters with attribute S (static) that are
not changed by the process. This parameter will be incremen‐
ted at least by one if at least one static parameter has been
modified. This provides a check of the parameter revision. In
case of an overflow this parameter is set to 1 i.e. 0 is not used.
33 IM_Profile_ID Unsigned16 Read only Fixed to 0x9700
34 IM_Profile_Specif‐
ic_Type
Unsigned16 Read only -
35 RESET Unsigned16 Read /
write
Valid Codes:
1: Application Reset:
Is the command for resetting a device to Profile
default values. The setting of the bus address is not affected.
2506: Restart
is the command for a restart of the device. All non-
volatile parameters re-main unchanged; all dynamic parame‐
ters are reset to their defaults.
2712: Communication Reset
The bus address is set to its default address; other
parameterization remains unchanged. The bus address is
changed immediately regardless if the device is in cyclic data
transfer state. The reset is not suspended up to a subsequent
power cycle / warm start. The No_Add_Chg_Flag correspond‐
ing to the Set_Slave_Add service is cleared.
2713: Application and Communication Reset
Combined Application Reset and Communication
Reset
36 IM_DESCRIPTOR Visible‐
String(54)
Read /
write
54 character long description field
37 LANGUAGE Visible‐
String(2)
Read /
write
Language setting of the local user interface. The coding is
according Alpha-2 of ISO639. Available codes:
“en”
“de”
“it”
“fr”:
“es”
“zh”
English
German
Italian
French
Spanish
Chinese
38 IM_Date Visible‐
String(16)
Read /
write
Installation date of the device.
39 STARTUP_PAR‐
AM_VALIDITY
Unsigned8 Read only Provides information if the cyclic master has sent startup pa‐
rameters that has been declared as valid to be taken to adjust
the device parameterization.
Communication
2.10 Acyclic communication
FCT030 Profibus (From firmware 4.0)
28 Function Manual, 06/2018, A5E39931906-AB