Loading...
Direct sunlight
Prevent the device from overheating or materials becoming brittle due to UV exposure by
protecting it from direct sunlight. Make sure that the maximum permissible ambient
temperature is not exceeded. Refer to the information in Operating conditions (Page 58).
WARNING
Equipment used in hazardous areas
Risk of explosion in hazardous areas.
Special requirements apply to the location and installation of the device. See Use in
hazardous areas (Page 19).
WARNING
Strong vibrations
Risk of explosion in hazardous areas.
In plants with strong vibrations, mount the transmitter in a low vibration environment.
CAUTION
Strong vibrations
Damage to device
In plants with strong vibrations, mount the transmitter in a low vibration environment away
from the sensor.
3.1.2 Proper mounting
NOTICE
Incorrect mounting
The device can be damaged, destroyed, or its functionality impaired through improper
mounting.
Before installing ensure there is no visible damage to the device.
Make sure that process connectors are clean, and suitable gaskets and glands are used.
Mount the device using suitable tools. Refer to the information in Technical data
(Page 57).
Installing/mounting
3.1 Basic safety notes
SITRANS FC430
Compact Operating Instructions, 09/2019, A5E39940293-AB 23

Table of Contents