6. Reinstate and tighten lid lock screw.
7. Ensure that moisture does not penetrate the inside of electronics enclosure. Bend cables
downward immediately before cable glands.
4.2.2.9 Finishing the transmitter connection (wall mount)
Connection check-up
1. Check individual wire installation by tugging firmly.
2. Firmly tighten cable glands and insert blanking plugs in unused cable entries.
3. Close lid.
4. Tighten the four spring screws.
5. Ensure that moisture does not penetrate to inside of electronics enclosure.