Service and maintenance
6
6.1 Basic safety notes
Note
The device is maintenance-free.
The device is maintenance-free. However, a periodic inspection according to pertinent
directives and regulations must be carried out.
An inspection can include, for example, check of:
Ambient conditions
Seal integrity of the process connections, cable entries, and cover
Reliability of power supply, lightning protection, and grounds
WARNING
Impermissible repair and maintenance of the device
Repair and maintenance must be carried out by Siemens authorized personnel only.
WARNING
Dust layers above 5 mm
Risk of explosion in hazardous areas.
Device may overheat due to dust build up.
Remove dust layers in excess of 5 mm.
NOTICE
Penetration of moisture into the device
Device damage
Make sure when carrying out cleaning and maintenance work that no moisture penetrates
the inside of the device.
SITRANS FC430
Compact Operating Instructions, 09/2019, A5E39940293-AB 51