7. Connect ground to terminal and power to terminals L/+ and N/- in the manner shown
below at right using a screwdriver.
Note
The terminal box is detachable
For easier access unplug the terminal box. After connecting the wires, plug the terminal box
back in.
L/+
N/-
Protective Earth (PE)
AC connection DC connection
/
1
/
1
Power: 100 to 240 V AC, 47 to 63 Hz
/
1
Power: 19.2 to 28.8 V DC
8. Tighten cable gland.
9. Mount power supply protection cover and fasten protection cover screw.
4.2.2.8 Finishing the transmitter connection (field mount)
Connection check-up
1. Check individual wire installation by tugging firmly.
2. Firmly tighten cable glands and insert blanking plugs in unused cable entries.
3. Remove O-ring from lid.
4. Reinstate lid and screw in until mechanical stop. Wind back lid by one turn.
5. Mount O-ring by pulling it over the lid and tighten lid cover until you feel friction from the O-
ring on both sides. Wind lid further by one quarter of a turn to seal on the O-ring.
Connecting
4.2 Connecting FC430
SITRANS FC430
46 Compact Operating Instructions, 09/2019, A5E39940293-AB