Take care when handling the cable and passing it through cable ducting that the plug is not
subjected to excessive tension (pulling) as the internal connections may be disengaged.
Note
Never pull the cable by the plug - only by the cable itself.
1. Connect DSL using the supplied 4-wire cable with M12 connectors.
Note
Grounding
The DSL cable screen is mechanically connected to the grounding terminal (PE), only when
the M12 plug is correctly tightened.
Wiring sensor and transmitter (sensor terminal compartment)
A: Prepare the cable by stripping it at both ends.
PP
/
PPPP
2UDQJH
<HOORZ
:KLWH
%OXH
Figure 4-1 Cable end
B: Connecting sensor terminal compartment
1. Remove lock screw and remove lid.
2. Remove one of the blind plugs and fit cable gland.
3. Remove cap and ferrule from cable gland and slide onto cable.
4. Push cable through open gland; anchor cable with clamp bar.
Connecting
4.2 Connecting FC430
SITRANS FC430
34 Compact Operating Instructions, 09/2019, A5E39940293-AB