What does Flow Limit mean on Siemens Measuring Instruments?
S
Sheri AllenSep 16, 2025
If your Siemens Measuring Instruments display a 'Flow Limit' alarm, it indicates a water network failure, such as a pipe burst. The alarm will remain active until the flow rate drops below the flow alarm limit.
S
Samuel AllenSep 17, 2025
What causes Reverse Flow Limit on Siemens Measuring Instruments?
G
Gilbert GoldenSep 17, 2025
If Siemens Measuring Instruments displays 'Reverse Flow Limit', it indicates a water network failure where the non-return valve is broken. The alarm remains active until the reverse flow rate drops below the reverse flow alarm limit.
F
fredcruzSep 18, 2025
Why is the flow direction wrong on my Siemens Measuring Instruments?
V
virginiarodriguezSep 18, 2025
If the flow rate appears correct, but the sign (flow direction) is wrong on your Siemens Measuring Instruments, first verify that the upstream sensor is connected to the upstream F-connector on the frontend (labeled A). Also, check the flow direction parameter in Setup > Sensor > Flow direction; it might be set to "negative" flow.
K
Kimberly HancockSep 18, 2025
Why does Siemens Measuring Instruments display active flow with large sound velocity error?
A
Allen WatkinsSep 18, 2025
If your Siemens Measuring Instruments display shows active flow with a large sound velocity error but no alarm codes, it might be due to incorrectly installed sensors or incorrect setup parameterization. Verify the pipe dimensions, sensor selection, mounting configuration, and spacing. Also, if the entered expected sound velocity is far from the actual fluid sound velocity, optimal performance won't be achieved. If possible, perform the sensor setup wizard again using the initially measured sound velocity for the expected sound velocity.
D
Daniel OdomSep 18, 2025
What to do if Siemens Measuring Instruments display shows no flow and measured sound velocity is 0 m/s with alarm codes 14, 15 and 34?
C
Cory TuckerSep 18, 2025
If your Siemens Measuring Instruments display shows no flow and the measured sound velocity is 0 m/s, and you're seeing alarm codes 14, 15, and 34, there could be several reasons. First, ensure the pipe is completely filled with liquid and consider relocating sensors to a normally full section, avoiding the top or bottom of horizontal pipes. Check the sensor cables for breaks or incorrect connections, verifying continuity by evaluating the DC resistance. Also, verify the pipe dimensions, sensor selection, mounting, and spacing, ensuring ultrasonic couplant is applied. Finally, the sensors or frontend hardware might be damaged; verify their operation using a pipe simulator 'Test block'.
Z
zmorrisSep 18, 2025
What to do if Siemens Measuring Instruments display is completely blank, but communication is still possible?
J
John StevensSep 18, 2025
If the display on your Siemens Measuring Instruments is completely blank, but communication is still possible on the service channel, check the cable connections between the display and Transmitter module. The HMI module may also be defective, in which case you should contact Technical Support.
B
Beth EscobarSep 18, 2025
What to do if Siemens Measuring Instruments display shows no flow and measured sound velocity is 0 m/s with alarm code 34?
J
Jamie PenaSep 18, 2025
If your Siemens Measuring Instruments display shows no flow and the measured sound velocity is 0 m/s, accompanied by alarm code 34, it could be due to incorrectly installed sensors or incorrect setup parameterization. Verify the actual pipe dimensions against the entered values, and check the sensor selection, mounting configuration, and spacing. Also, an incorrect expected fluid sound velocity can cause issues, so verify that the entered expected sound velocity is correct for the application. The pipe might also be mostly empty, causing the meter to operate on pipe noise; ensure the pipe is completely filled with liquid or relocate sensors to a section that is normally full, avoiding installation on the top or bottom of horizontal pipes.
A
Autumn SimmonsSep 18, 2025
What to do if Siemens Measuring Instruments display is completely blank and communication with service channel is not possible?
B
barrettashleySep 18, 2025
If the display on your Siemens Measuring Instruments is completely blank and communication with the service channel is not possible, check the mains power connection. Verify that the PSU and transmitter modules are fully plugged into the backplane.
G
Gregory RomeroSep 22, 2025
What to do if the keypad is not responsive on Siemens Measuring Instruments?
K
Kara TylerSep 22, 2025
If the keypad on your Siemens Measuring Instruments is not responsive, check the tightness of the lid screws. Cycle the power and allow several seconds for auto recalibration of the keypad.
R
Robert RileySep 22, 2025
Why are the datalogger and backup not working on my Siemens Measuring Instruments?
P
Paul GlennSep 22, 2025
If the datalogger and backup are not working on your Siemens Measuring Instruments, and the error indicates a failed chkdsk, the SD card may have failed. Remove and try to repair the SD card in a PC. Replace the SD card if repair is unsuccessful. If the SD card is not installed, insert it and verify that the SensorFlash is recognized as being installed in menu 3.7.2.