SFIDK.PS.029.E5.02 / SW 2.01.0712
3.4 Mounting the cables into SITRANS FUS060 connection module
The SITRANS F US ultrasonic flowmeter system contain 2 or 4 cables in fixed length of 3, 15,
30, 60, 90 or 120 m (9.84 / 49.21 / 98.43 / 196.85 / 295.28 / 393.70 ft).
The cables to be connected to the transducers according to the schematics 3.5.1 or 3.5.2.
Check and then enter the cable length in the transmitter.
Make sure the cables are equal in length in order to avoid different signal delays in the
signal processing.
3.3 Connecting the sensor cables to the transmitter
When opening the lid you will have direct access
to checking the correct transducer cable con-
nections.
For connection of the 4 cables, just carefully
press the cable into the cable glands until the
"snap function" fixes the cable inside the
connection module. Make sure the cable is
mounted correctly by smoothly pulling the cable.
Carefully press the appropriate cables into the
connection module until fully connected into the
plug.
The cable glands are tightened on the threaded
bush until the cable is sealed tightly (IP65).