Eliminating process errors
Only two general groups of errors are shown in the display. “Process error” and “device error”.
Device errors describe hardware errors. Two main groups of errors are:
Measuring path disturbed The medium in the measuring line is not permeable to
sound, this applies both when the pipe is filled with gas or
is empty. The measured value on the display is set to zero.
Cable is broken or detached.
Flow measurement unreliable The measured values have a very high dispersion so that
errors in the measuring evaluation or disturbances in the
flow due to cavitation, twist or inhomogenity such as bubbles
or foreign bodies.
Measured values are displayed but an “F” flashes in the
right hand corner of the display.
Error/Symptoms Diagnosis Solution/Comment
Display shows 0 flow. Pipe not filled with liquid Make sure that the pipe is filled or
A flashing “F” appears in the at least the sound path is covered
display continuously/occa- with liquid.
sionally.
Device status (menu 2.1): Cable broken/detached Make sure that the transducer
“Measurement path disturbed” cable is connected in the terminal
housing of the transmitter.
Gain (menus 6.5.1.X) for the
active tracks > 250 Make sure that the screen and
center pin is connected inside the
transducer top.
Time-of-flight (menu 6.5.5.X Permanent disturbance inside If possible, make sure that the
and 6.5.6.X) for the active the pipe transducer window inside the
tracks is not stable pipe is not damaged or that no
permanent disturbance is present.
Error counters (menu 6.5.4.X)
not equal to 0
Flow not stable. A flashing Air bubbles or particles of any Make sure that the pipe line is
“F” appears in the display kind disturbing the measurement ventilated and that the concentra-
continuously/occasionally. tion of particles is limited to a level
Device status (menu 2.1): at which normal measurement with
“Unreliable flow value” the meter is possible.
Gain (menus 6.5.1.X) for the Cable partly connected Make sure that the transducer
active tracks < 250 but not cables are connected in the
stable transmitter and in the sensors
(SONO 3200 or SONO 3300).
Check the connection of the
cable screening.
Time-of-flight (menu 6.5.5.X Make sure that the screen and
and 6.5.6.X) for the active center pin is connected inside
tracks is not stable the transducer top.
Error counters (menu 6.5.4.X) Ultrasonic echoes Make sure that the entered track
not equal to 0 lengths correspond with the
measurement report.
Trouble shooting 7