SFIDK.PS.029.E5.02 / SW 2.01.0740
Signal type menu 4.3.2
The function of the relay is determined in this menu item:
Contact closes: Relay contact closes at
- Alarm message Selection in menu “4.3.1 Function”
- Flow in forward-direction
- Limit signal
Contact opens: Relay contact opens at
- Alarm message Selection in menu “4.3.1 Function”
- Flow in forward-direction
- Limit signal
The table explains relay logic.
No alarm Alarm message
Setting Reverse flow Forward flow
No limit signalling Limit signalling
Contact closes
Contact opens
Switching status of the relay
Digital output 2
The electronic fuse is tripped in the event of overloading. The recovery time of the fuse is a few
minutes.
Function, menu 4.3.1
The digital output 2 (relay contact) can be configured for reporting device status (device fault,
measuring error), flow direction or a limit for flow, US velocity, US amplitude.
Possible alarm states are as follows:
Alarm - device state status
Flow direction forward
Flow min - flow below lower limit
Flow max - flow above upper limit
Flow min/max - flow either under below limit or above upper limit
Min/max US velocity
Max US level
Min US level
Min/max US amplitude
Max US amplitude
Min US amplitude
No function
Note
The relay contact is open when “no function” is set.