Procedure
1. Tap on "START FULL VERIFICATION". A full veri󹪝cation consists of transmitter test, insulation
test and coil circuit test.
You can also start a partial veri󹪝cation by switching to the "Veri󹪝cation" tab. No
certi󹪝cates can be generated on partial veri󹪝cations.
A test report in CSV 󹪝le format can be saved locally, for partial as well as full veri󹪝cations.
2. Wait for the veri󹪝cation to 󹪝nish. During the veri󹪝cation process the 󹪞owmeter turns ON and
OFF multiple times. At the end of the veri󹪝cation the 󹪞owmeter turns OFF.
3. Tap on "DISPLAY RESULTS". Wait for the results to be displayed.
4. The veri󹪝cation result is displayed. Scroll down through the results and tap on "SAVE DEVICE
DATA".
Operating
7.3Starting a veri󹪝cation and generating a certi󹪝cate or report
SITRANS Veri󹪝cator
Operating Instructions, 03/2024, A5E51481698-AA 33