Connecting
5.1 Connection recommendations
SITRANS LVS100/200
Operating Instructions, 01/2019, A5E44839609-AA
21
Conduit systems for FM and CSA (Dust and Gas Hazardous Locations)
General requirements: In addition the regulations of the country must be observed. The used
flameproof seals and blanking elements must have an adequate type approval and a
temperature range of at least -40 to 80 °C (-40 to 176 °F). In addition, they shall be suitable
for the conditions and correctly installed. Where available, the provided original parts of the
manufacturer must be used.
5.1
Connection recommendations
Use a fuse for the signal output (max. 10 A).
Provide protection for relay contacts to protect the device against spikes if inductive loads
are connected.
5.2
Precautions
Before opening the lid, ensure there are no dust deposits around SITRANS LVS100/200,
and that the atmosphere around the instrument is settled.
Make sure the main voltage does not exceed the maximum voltage listed on the product
label.
Ensure that no more than 8 mm of each wire is stripped (to avoid danger of contact with
live parts).
Ensure the boots for protecting cable terminations are no longer than 8 mm (to avoid
danger of contact with live parts).
For installation of the 3-wire PNP electronics (Protection Class III), the insulation of the
connected power supply and output load must meet the relevant standards.