Connecting
5.3 Power supply specifications
SITRANS LVS100/200
26 Operating Instructions, 01/2019, A5E44839609-AA
5.3
Power supply specifications
Universal voltage (DPDT relay)
(LVS100)
AC:
terminal 1: L
terminal 2: N
19 to 230 V AC, ± 10 %, 50 to 60 Hz, 22 VA
DC:
terminal 1: +
terminal 2: -
19 to 40 V DC, ± 10 %, 2 W
Universal voltage (SPDT relay)
(LVS200)
AC:
terminal 1: L
terminal 2: N
19 to 230 V AC, + 10 %, 50 to 60 Hz, 8 VA
DC:
terminal 1: +
terminal 2: -
19 to 55 V DC, + 10 %, 1.5 W
Universal voltage (DPDT relay)
(LVS200)
AC:
terminal 1: L
terminal 2: N
19 to 230 V AC, + 10 %, 50 to 60 Hz, 18 VA
DC:
terminal 1: +
terminal 2: -
19 to 55 V DC, + 10 %, 2 W
3-wire PNP (LVS200)
DC:
terminal 1: +
terminal 2: -
18 to 50 V DC, + 10 %, 1.5 W
2-wire (LVS200)
AC:
terminal 1: L
terminal 2: N
19 to 230 V AC, + 10 %, 50 to 60 Hz, 1.5 VA
DC:
terminal 1: +
terminal 2: -
19 to 230 V DC, + 10 %, 1 W
NAMUR IEC 60947-5-6
(LVS200)
ca. 7 to 9 V DC
8/16 mA or 4 to 20 mA (LVS200)
DC:
terminal 1: +
terminal 2: -
12.5 to 36 V DC, + 0 %