10
General
During scanning the user normally will get “Real time”
reconstructed images in full image quality, if the “fast”
slice has been selected.
In some cases – this depends also on Scan range, Feed/
Rotation and Reconstruction increment – the Recon
icon on the chronicle will be labeled with “RT”. This indi-
cates the Real time display of images during scanning.
The Real time displayed image series has to be recon-
structed afterwards.
The following tables show you the possibilities of image
reconstruction in spiral and sequential scanning.
Slice Collimation and Slice Width for Spiral Mode
12*/16** x 0.75 mm 0.75, 1, 1.5, 2, 3, 4, 5, 6, 7,
8, 10 mm
12*/16** x 1.5 mm 2, 3, 4, 5, 6, 7, 8, 10 mm
Cardio Spiral Modes
12 x 0.75 mm 0.75, 1.0, 1.5, 2,3 mm
12 x 1.5 mm 2, 3, 4, 5 mm
Slice Collimation and Slice Width for
Sequence Mode
12 x 0.75 mm 0.75, 1.5, 3, 4.5, 9 mm
12 x 1.5 mm 1.5, 3, 4.5, 6, 9 mm
2 x 5 mm 5, 10 mm
Perfusion Modes
12 x 1.5 mm 6, 9, 18 mm
16 x 1.5 mm 12mm
ECG triggered Modes
12 x 0.75 mm 0.75, 1.5, 3 mm
12 x 1.5 mm 1.5, 3, 6 mm
UHR Modes
2 x 0.6 mm 0.6 mm
2 x 1 mm 1, 2 mm
* For all Spiral Modes using 0.42 sec or 0.75 sec
Rotation Time.
** For all Spiral Modes using a 0.5, 1.0 or 1.5 sec
Rotation Time (exception : Cardio Protocols).