38
Children
InnerEarSeq
Indications:
High Resolution Spiral mode for inner ear studies,
e. g. malformations of the inner ear, inflammatory
changes, pathologies of the mastoid process,
tumorous processes of the pyramids, post-traumatic
changes, etc.
InnerEarSeq
kV 120
Effective mAs *
Slice collimation 0.6 mm
Slice width 0.6 mm
Feed/Scan 1.0 mm
Rotation time 0.75 sec.
Kernel U80u
Image order cr-ca
* The mAs should be adjusted to the age of the child.