134
Spine
RoutineSeq
Indications:
Sequence mode for spine studies, e. g. prolapse,
degenerative changes, trauma, tumors etc.
RoutineSeq
kV 120
Effective mAs 200
Slice collimation 0.75 mm
Slice width 1.5 mm
Feed/Scan 9.0 mm
Rotation time 1.0 sec.
Kernel B30s
Cycle time 3.0 sec.
Image order cr-ca
CTDI
w
16.8 mGy
Effective dose Male: 0.4 – 0.5 mSv
Female: 0.7 1.5 mSv