Indications:
Sequence mode for spine studies, e. g. prolapse,
degenerative changes, trauma, tumors etc.
A range of 20 cm will be covered in 12.1 s.
SpineSpi
kV 120
Effective mAs 300
Slice collimation 1.5 mm
Slice width 3.0 mm
Feed/Rotation 13.5 mm
Rotation time 0.75 sec.
Kernel B20f
Increment 3.0 mm
Image order cr-ca
CTDI
w
21.6 mGy
Effective dose Male: 6.0 mSv
Female: 10.2 mSv