54
ThoraxSeqHR
Indications:
Sequence mode for High Resolution lung studies, e. g.
interstitial changes of the lung parenchyma, etc.
ThoraxSeqHR
kV 120
Effective mAs *
Slice collimation 1.0 mm
Slice width 1.0 mm
Feed/Scan 10.0 mm
Rotation time 0.75 sec.
Kernel B70s
Image order cr-ca
Tips
• Children with a body weight of more than 54 kg
should be examined with an adult protocol.
• If you want to acquire the patient at full inspiration
or full expiration, you should practice the breathing
with the patient a few times before beginning the
scan, so that reproducibility may be improved.
* The mAs should be adjusted to the body weight
of the child.
* The conversion factor for an 8-week-old,
and a scan range of 150 mm was used.
Body weight kV mAs CTDI
w
Effective dose
(mGy) (mSv)
< 35 kg 120 30 0.5 Male: 0.08*
Female: 0.10*
35 – 54 kg 120 65 5.5
Children