ExtremityUHRSpi
Indications:
Spiral mode for Ultra High Resolution bone studies,
e. g. tumors, post-traumatic changes, orthopedic
indications, etc.
ExtremityUHRSpi
kV 120
Effective mAs *
Slice collimation 0.75 mm
Slice width 0.75 mm
Feed/Rotation 9.0 mm
Rotation time 0.75 sec.
Kernel U80u
Increment 0.5 mm
Image order cr-ca
* The mAs should be adjusted to the body weight
of the child.
Children
Body weight kV mAs CTDI
w
(mGy)
< 15 kg 120 26 2.2
15 34 kg 120 40 3.4
35 54 kg 120 80 6.7