105
Thorax
ThoraxCombi 2
nd
reconstr.
kV 120
Effective mAs 100
Slice collimation 0.75 mm
Slice width 5.0 mm 1.0 mm
Feed/Rotation 15.0 mm
Rotation time 0.5 sec.
Kernel B30f B80f
Increment 5.0 mm 0.7 mm
Image order cr-ca
CTDI
w
7.8 mGy
Effective dose Male: 3.0 mSv
Female: 3.9 mSv
If you want to reconstruct thin slices in every 2 mm
or 10 mm instead of 0.7 mm as predefined, simply
change the increment before image reconstruction.
MPRthick: ThoraxCombi coronal
Image thickness 3
Distance between images 3
Number of images 60
The 2
nd
reconstruction can be loaded in the 3D Card,
MPR. Select the Range function and a predefined range,
linked to the series, will pop up.
Please notice, if you are not satisfied with the Range
preset adapt the parameters to your needs and link them
to the series.
Contrast medium IV injection
Start delay 25 sec.
Flow rate 2.5 ml/sec.
Total amount 80 ml