Thorax
ThoraxRoutine
kV 120
Effective mAs 100
Slice collimation 1.5 mm
Slice width 6.0 mm
Feed/Rotation 30.0 mm
Rotation time 0.5 sec.
Kernel B40f
Increment 6.0 mm
Image order cr-ca
CTDI
w
7.0 mGy
Effective dose Male: 2.8 mSv
Female: 3.6 mSv
You could repeat the same protocol simply by clicking
the chronicle with the right mouse button for “repeat“.
E. g. when both non-contrast and contrast studies are
required.
For lung cancer staging, this protocol can be combined
with protocol “Neck Routine”.
Contrast medium IV injection
Start delay 25 30 sec.
Flow rate 2.5 ml/sec.
Total amount 80 ml
107