User Guide CGM
Attachment D Technical Information 82
D.6. Guidance and Manufacturer’s Declaration –
Electromagnetic Immunity
The transmitter is intended for use in the electromagnetic environment
specied in the next table. The customer or the user of the transmitter
should ensure that it is used in such an environment.
Immunity Test Transmitter Compliance Level
Electrostatic Discharge (ESD) IEC 61000-4-2
± 8 kV Contact
± 15 kV Air
Magnetic Field (50Hz and 60Hz) IEC 61000-4-8
30 A/m
Radiated Fields Disturbance IEC 61000-4-3
10 V/m at 80 MHz to 2700 MHz (AM Modulation)
Electromagnetic interference can still occur in the home healthcare
environment as control over the EMC environment cannot be
guaranteed. An interference event can be recognized by gaps in CGM
results or gross inaccuracies. The user is encouraged to try to mitigate
these eects by one of the following measures:
If your symptoms do not match your CGM results, use your blood
glucose meter when making treatment decisions. If your CGM results
do not consistently match your symptoms or blood glucose meter
values, then talk to your healthcare professional about how you
should be using the iCan i3 CGM to help manage your diabetes. Your
healthcare professional can help you decide how you should best use
this device.
D.7. iCan ACCESS and iCan REACH APP safety
statements
iCan ACCESS lets you send your sensor information from your App to
your Care Partners’ smart devices (iCan REACH APP). iCan REACH
APP’s information is always older than your App. The information on
iCan REACH APP is not meant to be used for treatment decisions or
analysis.