Chapter 5 Display and setting of various functions
5.19 Communication function HRS-R Series
5-66
Contact output signal 2 operation Setting and checking
53. Press the [SEL] key once.
The set screen of contact output signal 2 operation is displayed on the digital display.
54. Select contact output signal 2 operation from the table below with [▲] key or [▼] key,
and confirm by pressing “SEL”.
Table 5.19-28 List of set value
Set value
Explanation
Initial value
(Default setting)
N.O type
N.C type
Digital ouput signal 2 selected alarm Setting and checking
55. Press the [SEL] key once.
The set screen of contact output signal 2 selected alarm is displayed on the digital
display.
56. Select contact output signal 2 selected alarm from the table below with [▲] key or [▼]
key, and confirm by pressing “SEL”.
Table 5.19-29 List of set value
Set value
Explanation
Initial value
(Default setting)
Setting/checking are not available unless the
function setting of digial output 2 is selected
alarm status signal.
to
Sets selected alarm
Set range is AL.01 to AL.36.
Contact output signal 3 function Setting and checking
57. Press the [SEL] key once.
The set screen of contact output signal 3 function is displayed on the digital display.
PV
SV
PV
SV
PV
SV