How to correct operating troubles
Error messages Cause/Correction
Error codes: 2132 to 2134
Printing not possible: printer cannot be configured!
Check the settings in the menu Printer!
Ask a higher level user to check the settings of
the print function of the X-ray unit.
Error code: 2135
The image section selected with REVIEW is not print-
able anymore!
In case of insufficient memory capacity of the
electronics, older X-ray images can be displayed
by means of REVIEW, however, they cannot be
printed any more.
Error code: 2139
Printing not possible: no printable image!
The print key was actuated without any image be-
ing displayed.
Error code: 2140 see error code 2107
Error code: 2200
UPS unit does not respond!
The UPS unit of the power fail buffer* is defective
or not connected. If a power failure arises before
repairing the UPS unit, the automatic switching
off of the X-ray inspection unit is no longer guar-
anteed. If the power failure lasts more than 2
minutes, switch off the unit yourself by means of
the key switch.
Error code: 2201
Power failure or unit disconnected from mains!
See paragraph "Power failure" (F 1-94).
Error code: 2202
Batteries of UPS unit exhausted!
Even if you do not use the X-ray inspection unit, it
should always be connected to the mains to en-
sure that the buffer batteries are charged! If this
is not possible, you have to switch off the incor-
porated UPS unit, if accessible (F 1-31 - "How to
switch on"), or you have to have it switched off by
a competent technician.
Error codes: 2301 to 2303
Card is not correctly inserted or damaged!
If the error occurs although the card has been
correctly inserted, have a new ISIS card issued.
Error code: 2402
The current ambient temperature of the electronics is
45/50 °C! Failures may occur at a temperature of
more than 50 °C!
Arrange for appropriate cooling in order to avoid
that the temperature rises above 50°C. If this is
not possible, turn off the unit.
Error code: 2802
Event data base full! Possibly delete unnecessary im-
age files!
Ask a higher level user to delete image files which
are not needed any more.
Error code: 3100
Last Xport image transfer still in progress!
The error message and the conveyor stops in-
volved can be avoided by enlarging the distances
between the baggage and freight items.