Introduction 7828 Density Transmitter Technical Manual
4-8 78285000_AB
To exit from any of the configuration windows without making any changes, press the
Escape key on your computer.
Shows which unit is
being configured.
Enter values of
variable to give 4mA
and 20mA analog
outputs
Select variable to
control 4-20mA
output.
The calculated
parameters
(special
functions)
available depend
on whether
Matrix or API
referral is
selected.
Select units for
variable controlling
analog output
Enter line
pressure value
here.
Select from 1s to
100s, or no
averaging.
Select referral
type - see text for
more details.
Click on ‘Configure..’ to
select and configure
Special function (calculated
parameter). See text for
more details.
Select any combination of:
- General system fault
- Analog fault
- User selectable range
to invoke an alarm (2mA)
condition on the analog output.