Modbus Communications 7828 Density Transmitter Technical Manual
D-10 78285000_AA
D.5.8 REFERRAL TEMPERATURE
Used in Register 27
Index Referral Temperature
0 Lowest temperature value in matrix
1
2
3
4 Highest temperature value in matrix
D.5.9 ALARM COVERAGE
Used in Register 29.
Bit Pattern Coverage
0x00000001 4 - 20 mA output 1 alarm
0x00000008 System error
0x00000010 User defined alarm
D.5.10 ALARM HYSTERESIS
Used in Register 31.
Index 4 - 20 mA Output
Hysteresis
00 %
1 0.5 %
21 %
32 %
45 %
5 10 %
D.5.11 SOFTWARE VERSION
Used in Register 49.
Index Density Referral
0 Matrix
1 API

Table of Contents