7828 Density Transmitter Technical Manual Introduction
78285000_AB 4-11
4.4.6 REGISTER DUMP/LOAD
This facility is essential for saving the configuration of your 7828. You should use it to
save the current configuration before you start to alter it, in order to restore it if things go
wrong for any reason. Also, if you send the transmitter away for servicing or re-
calibration, you should save the current configuration.
Details are given below.
Address of unit
being accessed
Enter desired filename
for Dump, or required
filename for Load.
Choose data delimiter
(Dump only)
Choose which
sets of
registers to
save to file, or
simply save all
of them.
You can also
specify
individual
registers.
Restore a
previously
saved set of
register data
from file.
Store the
selected
register data
to a file.