EasyManua.ls Logo

Sony HCD-MD373 - Adjustment Procedures; Focus Bias Adjustment

Sony HCD-MD373
98 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
37
9. Press the ENTER/YES “R” button, and save the adjustment
results in the non-volatile memory. (“EFB = SAVE” will be
displayed for a moment.)
10. “EFB = MO-P”. will be displayed.
The optical pick-up moves to the pit area automatically and
servo is imposed.
11. Press the =0/MD/CD/TUNING – button or )+/
MD/CD/TUNING + button until the waveform of the oscillo-
scope moves closer to the specified value.
In this adjustment, waveform varies at intervals of approx. 2%.
Adjust the waveform so that the specified value is satisfied as
much as possible.
(Traverse Waveform)
12. Press the ENTER/YES “R” button, and save the adjustment
results in the non-volatile memory. (“EFB =
SAVE” will be
displayed for a moment.)
Next “EF MO ADJUST” is displayed. The disc stops rotating
automatically.
13. Press the 6 (MD) button and remove the disc.
14. Load the check disc (MD) TDYS-1.
15. Press the =0/MD/CD/TUNING – button or )+/
MD/CD/TUNING + button and display “EF CD ADJUST”.
16. Press the ENTER/YES “R” button and display “EFB =
CD”. Servo is imposed automatically.
17. Press the =0/MD/CD/TUNING – button or )+/
MD/CD/TUNING + button so that the waveform of the oscil-
loscope moves closer to the specified value.
In this adjustment, waveform varies at intervals of approx. 2%.
Adjust the waveform so that the specified value is satisfied as
much as possible.
(Traverse Waveform)
18. Press the ENTER/YES “R” button, display “EFB = SAVE”
for a moment and save the adjustment results in the non-vola-
tile memory.
Next “EF CD ADJUST” will be displayed.
19. Press the 6 (MD) button and remove the check disc (MD)
TDYS-1.
Note 1 : MO reading data will be erased during if a recorded disc is
used in this adjustment.
Note 2 : If the traverse waveform is not clear, connect the oscillo-
scope as shown in the following figure so that it can be
seen more clearly.
5-12. FOCUS BIAS ADJUSTMENT
Adjusting Procedure :
1. Load a test disk (MDW-74/AU-1).
2. Press the =0/MD/CD/TUNING – button or )+/
MD/CD/TUNING + button and display “CPLAY MODE”.
3. Press the ENTER/YES “R” button and display “CPLAY MID”.
4. Press the MENU/NO “R” button when “C1 = AD =
is displayed.
5. Press the =0/MD/CD/TUNING – button or )+/
MD/CD/TUNING + button and display “FBIAS ADJUS”.
6. Press the ENTER/YES “R” button and display “ / a =
”.
The first four digits indicate the C1 error rate, the two digits
after [/] indicate ADER, and the 2 digits after [a =] indicate the
focus bias value.
7. Press the )+/MD/CD/TUNING + button and find the
focus bias value at which the C1 error rate becomes about 200
(Refer to Note 2).
8. Press the ENTER/YES “R” button and display “ / b =
”.
9. Press the =0/MD/CD/TUNING – button and find the
focus bias value at which the C1 error rate becomes about 200.
10. Press the ENTER/YES “R” button and display “ / c =
”.
11. Check that the C1 error rate is below 50 and ADER is 00. Then
press the ENTER/YES “R” button.
12. If the “( ” in “ - - ( ” is above 20, press the ENTER/
YES “R” button.
If below 20, press the MENU/NO“R” button and repeat the
adjustment from step 2.
13. Press the 6 (MD) button to remove the test disc.
Note 1 : The relation between the C1 error and focus bias is as
shown in the following figure. Find points a and b in the
following figure using the above adjustment. The focal
point position C is automatically calculated from points a
and b.
Note 2 : As the C1 error rate changes, perform the adjustment us-
ing the average vale.
VC
A
B
Specification A = B
VC
A
B
Specification A = B
330 k
Oscilloscop
e
10pF
BD (MD) board
CN110 pin
3
(TE)
CN110 pin
1
(VC)
C1 error
about
200
b
c a Focus bias value
(F. BIAS)

Table of Contents

Related product manuals