Single and Multichannel Detector (SCD) / (MCD)
(21)
Therefore, from a spectral ratio plot the dead time can be determined. The detection
efficiency N’1 / N1 can now be calculated
N '
1
/
N
1
=
N '
1
/
N '
2
×1−
N '
2
/
A
(22)
Figure 17: Linearity Plot for the new Extended Range CEM
The count rate N’ was measured for different beam currents I. From the low count rate
region, the conversion factor c (N = c× I) can be calculated.