Chapter 6
Adjustment and Verification
P/N 79000035_SM, Version 1.5 6 - 55 Focus 30/35 Service Manual
Fig. 6-40 Instrument crosshair to measurement frame
5 Align Transmitter
beam to infinity.
A. Click Holding mode
ON.
B. Move the collimator
greyfilter to position
2,see Fig. 6-44 on
page 6-58 .
C. Switch Transmitter
ON and click Filter
motor
High,see Fig.
6-37 on page 6-52 .
D. Adjust half the
deviation of the spot
by adjusting the
vertical and
horizontal
adjustment screw,see
Fig. 6-43 on page 6-
58
E. Adjust the other half
by shifting the
transmitter,see Fig.
6-38 on page 6-53 .
1,5mm Allen
key
Switch Collimator
wavelength adapter to
the Pos. 2 = red.
Item Performance Description Result Notes