Chapter 6
Adjustment and Verification
Focus 30/35 Service Manual 6 - 82 P/N 79000035_SM, Version 1.5
Fig. 6-67 Set atmospheric conditions
7 Measure Baselines. A. Select
Offset
Determination
check
box.
B. Click
Start Offset
Determination
button,
see Fig. 6-67 on page
6-82 .
C. A message box with
the properties of the
first baseline appears.
D. Aim to the target and
click
Set Hz
orientation
.
E. Click
Measure to start
the distance
measurement, see
Fig. 6-68 on page 6-
83 .
F. Click
Position to
move instrument to
the next target.
Check correct
leveling of
instrument with
<10cc.
The measure
result and
standard
deviation is
shown after
each
measurement.
Selected baselines
will be displayed.
The shown tolerance
(500mm) on
distances for EDM
offset determination
(blue marked
distances) is for
correct distance
identification only.