Do not use the machine for
the following
Preparation of materials other than materials suitable for
materialographic studies. The machine must not be used for any
type of explosive and/or flammable material, or materials which
are not stable during machining, heating or pressure.
Models
Xmatic with high pressure cleaning and vertical conveyor
Xmatic with high pressure cleaning, ultra-sonic cleaning and vertical conveyor
Optional: Recirculation unit for MD grinding/polishing station.
2.2 Safety functions - (SF)
The machine is equipped with the following safety devices:
• Emergency stops - SF-1
Stops all hazardous movements
• Abrasive guard interlock - SF-2
Prevents the stone/disc from turning when the lid is open
• Limited speed function, stone - SF-3
Stops the motor if the rotational speed is exceeded
• Limited speed function, mover head - SF-4
Stops the motor if the rotational speed is exceeded
• Main safety cover interlock, hazardous movements - SF-5
Prevents any movement in the work zone when the main safety cover is open
• Main safety cover interlock, water and ethanol - SF-5A
Prevents water and ethanol from being dosed when the main safety cover is open
• Main safety cover interlock with locking device - SF-6
Prevents access to any moving part in case of after-run or loss of power
• MD elevator door locking device - SF-7
Prevents the elevator from moving when the door is open
• MD elevator door locking device - SF-8
Prevents access to the MD elevator in case of after-run or loss of power
• Vertical conveyor doors interlock - SF-9
Prevents movements in the vertical conveyor when the door is open
• Recirculation unit doors interlock, MD station - SF-10
Prevents the pumps from operating when the door is open
• Recirculation unit doors interlock, grinding stone station - SF-11
Prevents the pumps from operating when the door is open
• Alcohol exhaust timer - SF-12
Evacuates alcohol fumes from work zone area