Chapter 6 Operation
www.supmea.com
-40-
When selecting equivalent output, this parameter displays.
4 - Accumulation
4-1
Accumulation
clearance
Option
Manufacturer
Y, N
N
Clear accumulation amount.
4-2
Positive
accumulation
integer
Figure
Manufacturer
0-999999999
0
Set total positive integer part
4-3
Positive
accumulation
decimal
Figure
Manufacturer
0.0-0.999
0.0
Set total positive decimal part
4-4
Negative
accumulation
integer
Figure
Manufacturer
0-999999999
0
Set reverse total integer part
4-5
Negative
accumulation
decimal
Figure
Manufacturer
0.0-0.999
0.0
Set reverse total decimal part
5- Alarm contacts (3-0 set to show the configuration at alarm output )
No.
Type
Option
Password
level
Parameter
range
Default
5-0
Alarm 1
transistor
state
Option
User
High/Low
lever
High level
Touch spot outputs high and low level under no alarm state .
5-1
Alarm1 output
allowed
Option
User
Y/N
N
Allow contact 1 output of main switch, when it’s set to be N, the following
parameters do not display.
5-3
Allow alarm1
empty pipe
Option
User
Y/N
N