Chapter 6 Operation
www.supmea.com
-42-
7-3
Display alarm
permission
Option
User
Y/N
N
Allow alarm information to be displayed on the main screen
8- System
8-0
Language
Option
User
Chinese/
English
Chinese
Set configuration display language
8-1
Display
accuracy
Figure
User
0-4
2
The decimal digits of real-time volume.
8-2
Contrast
Figure
User
0-100%
50%
Contrast ratio of Liquid crystal display
8-3
Modbus
address
Figure
User
1-247
8
Communication Protocol instrument address based on the RS485
protocol Modbus RTU
8-4
Baud rate
Option
User
12002400
48009600
19200
38400
57600
9600
Baud rate of serial communication of physical layer
8-5
Even-odd
check
Option
User
NONE/ODD/
EVEN
NONE
Serial communication verification mode of physical layer
8-6
Byte swap
Option
User
2-1 4-33-4
1-24-3 1-2
1-2 3-4
2-1 4-3
Byte exchange sequence of physical layer serial communication
8-7
Device
address
Figure
User
0-999999
000001
HART device identification number