EasyManua.ls Logo

Swann AMI CACE - Page 44

Swann AMI CACE
82 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
42 A-96.250.871 / 120417
AMI CACE
Troubleshooting
E038 Temp. 2 Alarm low
check process
check programmed value, see
5.3.1.2.2.25, p. 64
E043 EDI out of range
check sample inlet pressure and
acknowledge this error message
If the problem persists, stop sample
flow and call service
E044 No sample flow
check sample inlet pressure.
check if the following components are
clogged:
inlet filter (if installed)
tubes
EDI module
If necessary, replace clogged parts.
See Tube numbering, p. 43 and
Replacing the EDI module, p. 44.
E045 EDI DAC disconnected
Stop sample flow and call service
E046 EDI ADC disconnected
Stop sample flow and call service
E049 Power-on
none, normal status
E050 Power-down
none, normal status
E065 EDI module exhausted
replace EDI module, see Replacing the
EDI module, p. 44.
Error Description Corrective action

Table of Contents

Related product manuals