42 A-96.250.871 / 120417
AMI CACE
Troubleshooting
E038 Temp. 2 Alarm low
– check process
– check programmed value, see
5.3.1.2.2.25, p. 64
E043 EDI out of range
– check sample inlet pressure and
acknowledge this error message
– If the problem persists, stop sample
flow and call service
E044 No sample flow
– check sample inlet pressure.
– check if the following components are
clogged:
inlet filter (if installed)
tubes
EDI module
– If necessary, replace clogged parts.
See Tube numbering, p. 43 and
Replacing the EDI module, p. 44.
E045 EDI DAC disconnected
– Stop sample flow and call service
E046 EDI ADC disconnected
– Stop sample flow and call service
E049 Power-on
– none, normal status
E050 Power-down
– none, normal status
E065 EDI module exhausted
– replace EDI module, see Replacing the
EDI module, p. 44.
Error Description Corrective action