Chapter 9 Troubleshooting
CS-5100 LAS Instructions for Use
Revised October 2014
9-3
9.2 Other information
The following errors appear in the [Detail] dialog box from the [Browser Main] screen.
(➤ CS-5100 Instructions for Use “Chapter 3: 3.9.3.2 Checking the details in case of abnormal analysis results”)
Code Message Cause Corrective actions/
countermeasures
24020 Sample transportation is
canceled by LAS
LAS sent a Sample ID Deletion
command.
Analysis was not completed and
must be repeated.
24021 Sample transportation is
canceled by LAS(All samples)
LAS sent a Sample ID Clear
command.
Analysis was not completed and
must be repeated.
24022 Sample aspiration is aborted by
LAS
LAS sent a Sample Aspiration
Abort command.
Analysis was not completed and
must be repeated.
24025 LAS communication is
disconnected.
1) The instrument or LAS
become power-off.
2) Communication error
Analysis was not completed and
must be repeated.