By default, the Peak/Valley method is turned off. To change test method to Peak/Valley using menu
system, (refer to section 3.3) get to the “nodE” (mode) options on Page 0 as shown below.
Using the “Digit Select” and “Digit Increment” buttons, set mode to desired peak / valley setup (refer
to menu tables in section 3.3.1 for setup explanations). Press the ‘Enter’ button to save.
5.4.3 Autosense/Continuous
The Autosense test method may be used when measuring many packages. Autosense avoids having
to press the ‘TEST’ button to initiate each test.
Autosense is a user-programmable parameter which specifies the delay time between each test in
seconds. When Autosense mode is selected for the analyzer and the green ‘TEST’ button is pressed,
the first package is tested as described for a timed test. The next test starts after the delay time specified
in “Auto” setup on Page 1. Upon completion of the first test the user has the specified delay time to
withdraw the needle from the first package, prepare and insert the needle into the next package, and
so on until all the packages have been measured.
To pause Autosense, press the green ‘TEST’ button or the ‘Enter’ button after the pump stops (before
next test starts). Press the green ‘TEST’ button again to re-start Autosense. To completely stop
Autosense, change “Auto” in Page 1 back to “0”.
To change Autosense to Continuous mode, change “Auto” in Page 1 to “1”. This will cause the pump
to run continuously until the green ‘TEST’ button is pressed to stop/start measurements or Autosense
is turned off as described above.
By default, Autosense is turned off. To change Test method to Autosense, using menu system (refer
to section 3.3) get to the “Auto” options on Page 1 of menus as shown below:
Using the ‘Digit Select’ and ‘Digit Increment’ buttons, change “Auto” from “0” to the desired time in
seconds between consecutive tests. Press the ‘Enter’ button to save. To turn Autosense method off
change “Auto” time back to “0”.