Gaspace 6000 Operations Manual v1.4
www.systechillinois.com
Page
11
Sets the password to access the Page 2 menu.
Sets the percent concentration of oxygen for calibration using certified gas (1% to 10% or
30% to 100%).
Sets the time in 24-hour format (HHMM).
Sets the date in format DDMM.
Sets the year four-digit format (YYYY).
Selects the menu page. Page 0 should be selected to return to the measurement
mode. Page 2 parameters are password protected. The factory default password is
1234.
3.3.3 GS6700 Menu System
When navigating through the menus, the display will flash the parameter name followed by its value
.
DISPLAYED PARAMETER
FUNCTION
Page 0
Selects which product code and its corresponding description will be printed. The user may
enter up to 9 product codes (1 through 9).
Selects the menu page. Page 1 or 2 parameters may be password protected. The
factory default password is 1234.
Page 1 (optionally password protected. Default password is 1234)
Sets the duration of a timed test in seconds (1 to 200).
Sets the delay time between tests for Autosense in seconds. 0=disable Autosense,
1=Continuous mode, >1=delay time up to 200 seconds.
Sets the Peak/Valley measurement mode for CO
2
/O
2
with the first digit corresponding to
CO
2
and the second digit corresponding to O
2
. 0=disable Peak/Valley, 1=Peak, 2=Valley i.e.
12” will set the Peak for CO
2
and Valley for O
2
.
Sets the alarm option.
0=alarm disabled, 1=alarm enabled.
Sets the value at which the carbon dioxide alarm will be triggered (0 to 100). This will not be
shown if the alarms are disabled (“ALEn” or Alarm Enable set to “0”).
Sets the carbon dioxide alarm polarity. 0=low alarm, 1=high alarm. This will not be shown if
the alarms are disabled (“ALEn” or Alarm Enable set to “0”).
Selects page (Page 0, 1). Page 1 parameters are password protected. The factory
default password is 1234.