Performance Checklcalibration-314 Service
TABLE
5-1
(cont)
Examples of Applicable
Test Equipment
a. Triplett Model 630NA
VOM.
Description
DC voltmeter1
Minimum Specifications
Range: 0 to 2000 V;
Accuracy, checked to within
1% at 1900 volts.
Usage
High Voltage supply check
and adjustment.
b. Simpson Model 262
VOM.
a. Trygon Power Supply
Model HR40-5B (0-40 V
at 5 A).
Adjust Regulator turnoff.
Check Low Line indicator.
Metered Variable
DC Supply1
Range:
0 to 28 V at 2 A.
Time Mark Generator
Markers from 0.1 to 5
s;
Marker accuracy, within 1%.
Vertical Geometry, Hori-
zontal Timing check and
adjustment.
a. Tektronix TG 501'.
b. Tektronix 2901.
c. Tektronix 184.
Standard Amplitude
Cali
brator
Range: 5 mV to 50 V;
Accuracy, within 0.25% of
amplitude setting into
1
Mf2 load.
Adjust and check gain.
Check
Volts/Div accuracy.
a. Tektronix 067-0502-01
Calibration Fixture.
b. Tektronix PG 506
Calibration Generator2.
Square-Wave
Generator
Repetition Rate: 500 kHz;
Risetime of leading edge
of positive pulse, 1 ns.
Check and Adjust
Volts/
Div compensation. Check
and adjust vertical high
frequency compensation.
a. Tektronix 106 Square-
Wave Generator.
b. Tektronix PG 506
Calibration
ene era to?.
a. Tektronix 191 Constant
Amplitude Signal
Generator.
Frequency Range: to at least
15 MHz; reference frequency,
50 kHz; Output amplitude, to
5 V (p-p), adjustable, into
50 f2; amplitude accuracy,
within
3% of reference fre-
quency amplitude as output
frequency changes.
Constant Amplitude
Signal Generator
Check vertical bandwidth.
Function Generator Frequency Range: 0.5 Hz to
1 MHz.
Check low-frequency
triggering.
a. Tektronix FG 501'
Cable (2 required) Impedance, 50 Q; connect-
ors, BNC; length 42 inches.
Impedance, 50 Q; connect-
ors, BNC.
Used throughout procedures
for signal interconnection.
Used throughout procedures
to terminate signal source.
Tektronix part number
Dl
2-0057-01.
Tektronix part number
01 1-0049-01.
Termination
Feedthrough
'use with Adjustment procedure only.
2~se with TM
500
Series Power Module.
REV. A, AUG.
1975