Section 3: Instrument description Model 2601B-PULSE System SourceMeter® Instrument User's Manual
3-12 2601B-PULSE-900-01 Rev. A August 2021
Descriptions of the configuration menus
The following table contains descriptions of the configuration menus and cross-references to
related information.
To access, press the
CONFIG key and then:
For more information, see:
V-source sense and low range;
I-source low range; and
high capacitance mode
Model 2601B-PULSE Reference Manual,
“Range”
Basic source-measure procedure
V and I-measure range,
V-measure sense, low range, and
autozero
Model 2601B-PULSE Reference Manual,
“Range”
Basic source-measure procedure
V-source and I-source compliance
limits
Model 2601B-PULSE Reference Manual,
“Speed”
Model 2601B-PULSE Reference Manual,
“Relative offset”
Model 2601B-PULSE Reference Manual,
“Filters”
Output-off states (on page 4-21)
Using the interlock (on page 2-37)
Set buffer count and destination
Model 2601B-PULSE Reference Manual,
“Source-measure concepts”
Set trigger in, count, interval,
and delay
Model 2601B-PULSE Reference Manual,
“Triggering”
Setting values
Through the front panel, you can adjust a value using either the Navigation wheel method or
Numeric entry method (using the keypad).
When the pulser is enabled, you cannot adjust values through the front panel.
Setting a value
Navigation wheel method:
1. Use the CURSOR arrow keys (or turn the navigation wheel) to move the cursor to the digit that
needs to be changed.
2. Press the navigation wheel or the ENTER key to enter edit mode. The EDIT indicator is
illuminated.
3. Rotate the navigation wheel to set the appropriate value.
4. Press the ENTER key to select the value or press the EXIT (LOCAL) key to cancel the change.
5. To return to the main menu, press the EXIT (LOCAL) key.