Section 3: Instrument description Model 2601B-PULSE System SourceMeter® Instrument User's Manual
3-26 2601B-PULSE-900-01 Rev. A August 2021
Adding USER TESTS menu entries
You can use the following function in either of two ways to add an entry into the USER TESTS menu:
display.loadmenu.add(displayname, code)
display.loadmenu.add(displayname, code, memory)
Where:
displayname
The name string that is added to the USER TESTS menu.
code
The code that is run from the USER TESTS menu when the RUN button is pressed.
It can include any valid Lua code.
memory
A value that specifies if the code and displayname parameters are saved in
nonvolatile memory. Set to one of the following values:
0 or display.DONT_SAVE
1 or display.SAVE (this is the default setting)
Scripts, functions, and variables that are used in the code are not saved when display.SAVE is
used. Functions and variables need to be saved with the script. If the script is not saved in nonvolatile
memory, it is lost when the 2601B-PULSE is turned off. See Example 1 below.
Example 1:
Assume a script with a function named DUT1 has been loaded into the 2601B-PULSE, and the script
has not been saved in nonvolatile memory.
Now assume you want to add a test named Test to the USER TESTS menu. You want the test to
run the function named DUT1 and sound the beeper. The following programming example illustrates
how to add Test to the menu, define the code, and then save displayname and code in
nonvolatile memory:
display.loadmenu.add("Test", "DUT1() beeper.beep(2, 500)", display.SAVE)
When Test is run from the front-panel USER TESTS menu, the function named DUT1 executes and
the beeper beeps for two seconds.
Now assume you turn the 2601B-PULSE power off and then on again. Because the script was not
saved in nonvolatile memory, the function named DUT1 is lost. When Test is again run from the front
panel, the beeper beeps, but DUT1 does not execute because it is no longer in the
runtime environment.
Example 2:
The following command adds an entry called Part1 to the front-panel USER TESTS submenu for the
code testpart([[Part1]], 5.0) and saves it in nonvolatile memory:
display.loadmenu.add("Part1", "testpart([[Part1]], 5.0)", display.SAVE)