Commands Listed in Alphabetical Order
Examples
ACQUIRE:NUMAV
G16
specifies that 16 waveform averages will be performed
before exponential averaging starts.
ACQUIRE:NUMA
VG?
might return :ACQUIRE:NUMAVG 64 indicating that there
are 64 acquisitions specified for averaging.
ACQuire:NUMEnv
This command controls the number of envelopes (when acquisition mode has
been set to ENVelope using ACQuire:MODe). The number of envelopes can
be set from 1 to 2000 in increments of 1, or to INFInite. Setting the value to a
number gr
eater than 2000 sets the number of envelopes to INFInite.
Group
Acquisition
Syntax
ACQuire:NUMEnv {<NR1>|INFInite}
ACQuire:NUMEnv?
Related Commands
ACQuire:MODe, ACQuire:STATE
Arguments
<NR1> is an integer that specifies the number of envelopes to use when the
acquisition mode has been set to ENVelope.
INFInite specifies to use an infinite number of envelopes.
Examples
ACQuire:NUMEnv 22 sets the number of envelopes to 22.
ACQuire:NUMEnv? returns the number of envelopes that has been specified,
either an integer or INFINITE.
ACQuire:STATE
Starts or stops acquisitions. When state is set to ON or RUN, a new acquisition
will be started. If the last acquisition was a single acquisition sequence,anew
single sequence acquisition will be started. If the last acquisition was continuous,
a new continuous acquisition will be started.
If RUN is issued in the process of completing a single sequence acquisition (for
example, averaging or enveloping), the acquisition sequence is restarted, and
any accumulated data is discarded. Also, the oscilloscope resets the number of
acquisitions. If the RUN argument
is issued while in continuous mode, acquisition
continues.
2-116 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual