Commands Listed in Alphabetical Order
Group
Acquisition
Syntax
ACQuire:STATE {OFF|ON|RUN|STOP|<NR1>}
ACQuire:STATE?
Related Commands
ACQuire:STOPAfter
Arguments
OFF stops acquisitions.
STOP stops acquisitions.
ON starts acquisitions.
RUN starts acquisitions.
<NR1> = 0 stops acquisitions; any other value starts acquisitions.
Examples
ACQUIRE:STATE RUN starts the acquisition of waveform data and resets the count
of the number of acquisitions.
ACQUIRE:STATE? might return:A CQUIRE:STATE 0 indicating that the
acquisition is stopped.
ACQuire:STOPAfter
Specifies whether the oscilloscope should continually acquire acquisitions or
acquire only a single sequence.
Group
Acquisition
Syntax
ACQuire:STOPAfter {RUNSTop|SEQuence}
ACQuire:STOPAfter?
Related Commands
ACQ
uire:STATE
Arguments
RUNSTop specifies that the oscilloscope will continually acquire data, if
ACQuire:STATE is turned on.
SEQuence specifies that the next acquisition will be a single-sequence acquisition.
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-117