Commands Listed in Alphabetical Order
BUS:B<x>:RS2
32C:DELIMiter
This command specifies the delimiting value for a packet on the RS-232 bus.
Conditions
This command requires a DPO4COMP application module for 4000 series
models, and an MDO3COMP application module for MDO3000 series models.
Group
Bus
Syntax
BUS:B<x>:RS232C:DELIMiter {NULl|LF|CR|SPace|XFF}
BUS:B<x>:RS232C:DELIMiter?
Arguments
NULl specifies 0x00.
LF specifies 0x0A.
CR specifies 0x0D.
XFF specifies 0xFF.
BUS:B<x>:RS232C:DISplaymode
This command specifies the display mode for the RS-232 bus (frame or packet).
Conditions
This command requires a DPO4COMP application module for 4000 series
models, and an MDO3COMP application module for MDO3000 series models.
Group
Bus
Syntax
BUS:B<x>:RS232C:DISplaymode {FRAme|PACKET}
BUS:B<x>:RS232C:DISplaymode?
Arguments
FRAme displays each frame as a single entity.
PACKET displays a group of frames terminated with a single frame defined by the
BUS:B<x>:RS232C:DELImiter command or the front panel.
BUS:B<x>:RS232C:PARity
This command specifies the parity for the RS-232C bus.
2-192 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual