Commands Listed in Alphabetical Order
Group
Bus
Syntax
BUS:B<x>:SPI:BITOrder {LSB|MSB}
BUS:B<x>:SPI:BITOrder?
Arguments
LSB sets the bit order to least significant bit first.
MSB sets the bit order to most significant bit first.
Examples
BUS:B1:SPI:BITORDER LSB sets the bit order to least significant bit first.
BUS:B1:SPI:BITORDER? might return BUS:B1:SPI:BITORDER MSB
indicating the bit order is set to most significant bit first.
BUS:B<x>:SPI{:CLOCk|:SCLk}:POLarity
This command specifies the SCLK polarity for the SPI bus.
Conditions
This command requires a DPO4EMBD application module for 4000 series
models, and an MDO3EMBD application module for MDO3000 series models.
Group
Bus
Syntax
BUS:B<x>:SPI{:CLOCk|:SCLk}:POLarity {FALL|RISe}
BUS:B<x>:SPI{:CLOCk|:SCLk}:POLarity?
Arguments
FALL specifies the SCLK polarity as falling edge.
RISe specifies the SCLK polarity as rising edge.
BUS:B<x>:SPI{:CLOCk|:SCLk}:SOUrce
This command specifies the SCLK source waveform for the SPI bus.
Conditions
This command requires a DPO4EMBD application module for 4000 series
models, and an MDO3EMBD application module for MDO3000 series models.
Group
Bus
MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual 2-195

Table of Contents

Other manuals for Tektronix MDO4000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix MDO4000B Series and is the answer not in the manual?

Tektronix MDO4000B Series Specifications

General IconGeneral
BrandTektronix
ModelMDO4000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals