Commands Listed in Alphabetical Order
Group
Histogram
Syntax
HIStogram:DISplay {OFF|LOG|LINEAr}
HIStogram:DISplay?
Related Commands
HIStogram:MODe
Arguments
<LOG> sets the histogram display to logarithmic scaling
<LINEAr> sets the histogram display to linear scaling. This is the default setting.
Examples
HISTOGRAM:DISPLAYLOG sets the histogram scaling to be the log count of each
histogram bin.
HISTOGRAM:DISPLAY? might return :HISTOGRA M:DISPLAY LOG indicating
that the current histogram display scaling is the log count of each histogram bin.
HIStogram:END? (Query Only)
Returns the time (horizontal) or vertical units value (vertical) of the last bin where
the histogram data ends. The HIStogram:MODe must be either
HORizontal or
VERTical for a value to be returned. If the mode is OF F, an error event is set
and nothing is returned.
If the HIStogram:MODe is
HORizontal, the value returned is the time of the
right bin. If the HIStogram:MODe is
VERTical the value returned is the vertical
units value of the bottom bin. The returned value is an <NR3>.
Group
Histogram
Syntax
HIStogram:END
?
Related Commands
HIStogram:MODe
HIStogram:STARt?
Returns
<NR3>
Examples
HIS
TOGRAM:END?
might return :HISTOGRAM:END 1.6000E-05 indicating that
the last bin is at 16 μs.
2-342 MDO4000/B/C, MSO/DPO4000B and MDO3000 Series Oscilloscopes Programmer Manual